Calibration Device for Spring Impact Hammer

Original
price: Negotiable
minimum:
Total supply:
Delivery term: The date of payment from buyers deliver within days
seat: Beijing
Validity to: Long-term effective
Last update: 2017-12-23 02:45
Browse the number: 149
inquiry
Company Profile
 
 
Product details
Principle of calibration:

The principle of this calibration procedure is to compare the energy provided by a spring hammer, which is difficult to measure directly, to the energy of a pendulum, calculated from its mass and height of fall. And it's designed with the requirements of Annex B of IEC60068-2-75.



Construction:

It's maily made of a frame, a bearing, a drag pointer, a pendulum, a release base and a release device.



Notes:

The Calibration Device fo Spring Hammer can be used for all models, allowing verification of the proper functioning of different types of hammers with the same calibration.



The spring is designed for calibrating spring hammers having characteristics as defined for energy values equal to less than 2J.



Conforms to:

IEC60335, IEC60065, IEC60745, IEC61029, IEC60601, IEC60610 etc.



Technology parameters:


Características del Equipo


Pendulum specifications: •Material: Steel.
•Diameter: Ø 20mm.
•Length: 635mm ± 2,5mm.
•Pendulum H : 120mm ± 2mm.
•Pendulum D: 20mm ± 0,5mm.
•Pendulum W: 4mm ± 0,01mm.

Impact energy: 0 a 2J.
Weight: 21 Kg.
Release device: Ac. Mechanic.
Finish: •Chrome bathroom.
•Painting.

According to: •Annex B of IEC 60068-2-75 fig. B1